ETG 9500 P Anesthetic Gases Monitor – Portable Anesthetic Gas Analyzer
Portable anesthetic gas analyzer

Portable FTIR anesthetic gas analyzer for operating rooms

Portable FTIR anesthetic gas analyzer · battery-powered · tablet touchscreen · optional multipoint case · sevoflurane, isoflurane, desflurane, nitrous oxide

ETG 9500 P Anesthetic Gases Monitor portable anesthetic gas analyzer with tablet in rugged case

ETG 9500 P Anesthetic Gases Monitor is a portable FTIR analyzer for anesthetic gases, in a rugged case ready to measure where it is needed. In the standard configuration it samples directly and shows every configured compound on its own tablet touchscreen, with no supervision software.

It is battery-powered, with up to 10 hours of operation on a single charge. An optional multipoint sampling case extends the system to continuous coverage of several rooms at once.

Up to 10 h
Battery operation
36 VDC · charging 3-4 h
< 1 ppm
Detection capability
0.5 cm⁻¹ resolution
15 kg
Weight
+10 kg with multipoint case
1–5 L/min
Sample flow
Direct sampling
Key features

Built for spot checks and temporary campaigns

01

Battery-powered

36 VDC rechargeable battery, up to 10 hours on a single charge (charging time 3-4 hours): no mains outlet needed on site.
02

Tablet interface

Real-time list of every configured compound, 30-minute trend, battery level and residual time on the same screen. Data export by USB drive or over the network (SMB).
03

Multipoint case (optional)

A dedicated sampling case (10 kg) adds a multipoint manifold for continuous coverage of several rooms.
Where it is used

Applications of the ETG 9500 P Anesthetic Gases Monitor

  • Periodic exposure checks in operating rooms and recovery rooms
  • Immediate verification after a suspected leak or an equipment change
  • Spot checks during commissioning or inspections
  • Coverage of rooms without a fixed monitoring system

See the anesthetic gas monitoring application for details.

FTIR vs other techniques

FTIR advantages over NDIR, gas chromatography and photoacoustic systems

Anesthetic gas monitoring has historically relied on NDIR analyzers, gas chromatography (GC) and, more recently, photoacoustic spectroscopy (PAS) with an infrared source. NDIR and PAS systems are generally based on a dedicated optical filter for each measured gas; the FTIR analyzer acquires the entire infrared spectrum and derives the concentration of all compounds of interest at once.

vs NDIR

One analyzer, many gases

A single analyzer measures sevoflurane, isoflurane, desflurane, nitrous oxide, carbon dioxide and water vapor without a dedicated sensor or filter for each gas. The 0.5 cm⁻¹ resolution separates halogenated anesthetics with partially overlapping bands, reducing the cross-interference typical of filter-based systems.
vs gas chromatography

Faster, no carrier gas

Analysis times are much shorter than a gas chromatograph, which needs a separation cycle for each measurement. No Helium or Hydrogen carrier gas, and no columns to replace: lower running costs over time.
vs photoacoustic (PAS)

Constant time, no alcohol interference

Measurement time does not grow with the number of gases. The full spectrum recognizes isopropanol and other alcohols from disinfectants as such, instead of counting them as anesthetic gas: in optical-filter photoacoustic systems this interference is documented in the scientific literature (Herzog-Niescery et al., BMC Anesthesiology, 2019): sevoflurane peaks of about 15 ppm, falling below 1 ppm once the alcohol peaks are removed.
Technical data

ETG 9500 P Anesthetic Gases Monitor: key specifications

FeatureDetail
Power supplyRechargeable battery, 36 VDC, up to 10 h on a single charge (charging time 3-4 h); charging from mains
Power consumptionMax 165 W · typical 120 W
ConnectivityEthernet, Wi-Fi, USB
Dimensions (L × W × D)500 × 280 × 400 mm (standard single-point case)
Weight15 kg (standard single-point case); +10 kg with the optional multipoint sampling case
RatingIP54 (optional)
Sample gas flow1 to 5 L/min
Gas fittingsQuick fittings for 6 mm tubing
Multipoint case (optional)10 kg

Operating conditions

FeatureDetail
Sample gas pressureMaximum 1 bar (relative): the instrument draws the ambient air to be analyzed with its own pump
Sample gas temperature15 °C to 60 °C
Measuring cell temperature60 °C: the sample gas is brought to this temperature before analysis
Ambient conditionsIndoor use, up to 2000 m altitude. 5 °C to 40 °C; relative humidity 20% to 80% up to 31 °C, decreasing to 50% at 40 °C, non-condensing
Warm-up time30 minutes

Analytical performance

FeatureDetail
Zero drift< 1% F.S. for every 2 weeks elapsed since the background
Linearity deviation< 1% F.S.
Repeatability≤ 0.5% of the measured value
Temperature-induced drift< 1% F.S. per 10 K of temperature change; ambient temperature changes are measured and compensated
Pressure influence1% F.S. for every 1% change in sample gas pressure; ambient pressure changes are measured and compensated
Response time (T90)30 seconds per sampling point
BackgroundEvery 2 weeks (recommended), with nitrogen 5.0 connected only during background acquisition: a permanent nitrogen line is not required
Full cycle timeT90 × number of configured points (single point in the standard configuration)

Measured gases

Measured gasRangeLDL (lower detection limit)
H2O0 – 5 %0.05 %
CO20 – 2000 ppm10 ppm
Sevoflurane0 – 20 ppm0.1 ppm
Isoflurane0 – 20 ppm0.1 ppm
Desflurane0 – 20 ppm0.1 ppm
N2O0 – 100 ppm0.25 ppm
Isopropanol0 – 100 ppm0.5 ppm
Ethanol0 – 100 ppm0.5 ppm

Options and accessories

Multipoint sampling caseIP54Additional gases on request
How it works

How FTIR spectroscopy works

With FTIR (Fourier transform infrared) spectroscopy every compound absorbs infrared radiation at specific frequencies: analyzing the whole spectrum gives the concentration of every gas, all at the same time and continuously.

The ETG 9500 FTIR analyzer has a 0.5 cm⁻¹ resolution and a 5-metre optical path in a low-volume cell, for high selectivity and sensitivity, with detection limits well below 1 ppm. Automatic background acquisition, internal filters, cleaning valves and temperature and pressure control keep performance stable over time.

ETG 9500 anesthetic gas monitors are measuring instruments for occupational exposure monitoring. They are not medical devices under Regulation (EU) 2017/745: they have no diagnostic or therapeutic purpose on the patient and are not intended for monitoring vital signs.
The ETG 9500 anesthetic gas monitors

Two systems, one FTIR platform

Cabinet

ETG 9500 Anesthetic Gases Monitor

Cabinet system with multipoint sampling for continuous monitoring of up to 48 points. Discover it →
Family

ETG 9500 anesthetic gas monitors

FTIR technology, continuous monitoring and the two systems at a glance. Read more →
Request information about the ETG 9500 P Anesthetic Gases Monitor
Tell us how you plan to use it: we will help you choose the configuration and options best suited.
Contact ETG