Portable FTIR anesthetic gas analyzer for operating rooms
Portable FTIR anesthetic gas analyzer · battery-powered · tablet touchscreen · optional multipoint case · sevoflurane, isoflurane, desflurane, nitrous oxide

ETG 9500 P Anesthetic Gases Monitor is a portable FTIR analyzer for anesthetic gases, in a rugged case ready to measure where it is needed. In the standard configuration it samples directly and shows every configured compound on its own tablet touchscreen, with no supervision software.
It is battery-powered, with up to 10 hours of operation on a single charge. An optional multipoint sampling case extends the system to continuous coverage of several rooms at once.
Built for spot checks and temporary campaigns
Battery-powered
Tablet interface
Multipoint case (optional)
Applications of the ETG 9500 P Anesthetic Gases Monitor
- Periodic exposure checks in operating rooms and recovery rooms
- Immediate verification after a suspected leak or an equipment change
- Spot checks during commissioning or inspections
- Coverage of rooms without a fixed monitoring system
See the anesthetic gas monitoring application for details.
FTIR advantages over NDIR, gas chromatography and photoacoustic systems
Anesthetic gas monitoring has historically relied on NDIR analyzers, gas chromatography (GC) and, more recently, photoacoustic spectroscopy (PAS) with an infrared source. NDIR and PAS systems are generally based on a dedicated optical filter for each measured gas; the FTIR analyzer acquires the entire infrared spectrum and derives the concentration of all compounds of interest at once.
One analyzer, many gases
Faster, no carrier gas
Constant time, no alcohol interference
ETG 9500 P Anesthetic Gases Monitor: key specifications
| Feature | Detail |
|---|---|
| Power supply | Rechargeable battery, 36 VDC, up to 10 h on a single charge (charging time 3-4 h); charging from mains |
| Power consumption | Max 165 W · typical 120 W |
| Connectivity | Ethernet, Wi-Fi, USB |
| Dimensions (L × W × D) | 500 × 280 × 400 mm (standard single-point case) |
| Weight | 15 kg (standard single-point case); +10 kg with the optional multipoint sampling case |
| Rating | IP54 (optional) |
| Sample gas flow | 1 to 5 L/min |
| Gas fittings | Quick fittings for 6 mm tubing |
| Multipoint case (optional) | 10 kg |
Operating conditions
| Feature | Detail |
|---|---|
| Sample gas pressure | Maximum 1 bar (relative): the instrument draws the ambient air to be analyzed with its own pump |
| Sample gas temperature | 15 °C to 60 °C |
| Measuring cell temperature | 60 °C: the sample gas is brought to this temperature before analysis |
| Ambient conditions | Indoor use, up to 2000 m altitude. 5 °C to 40 °C; relative humidity 20% to 80% up to 31 °C, decreasing to 50% at 40 °C, non-condensing |
| Warm-up time | 30 minutes |
Analytical performance
| Feature | Detail |
|---|---|
| Zero drift | < 1% F.S. for every 2 weeks elapsed since the background |
| Linearity deviation | < 1% F.S. |
| Repeatability | ≤ 0.5% of the measured value |
| Temperature-induced drift | < 1% F.S. per 10 K of temperature change; ambient temperature changes are measured and compensated |
| Pressure influence | 1% F.S. for every 1% change in sample gas pressure; ambient pressure changes are measured and compensated |
| Response time (T90) | 30 seconds per sampling point |
| Background | Every 2 weeks (recommended), with nitrogen 5.0 connected only during background acquisition: a permanent nitrogen line is not required |
| Full cycle time | T90 × number of configured points (single point in the standard configuration) |
Measured gases
| Measured gas | Range | LDL (lower detection limit) |
|---|---|---|
| H2O | 0 – 5 % | 0.05 % |
| CO2 | 0 – 2000 ppm | 10 ppm |
| Sevoflurane | 0 – 20 ppm | 0.1 ppm |
| Isoflurane | 0 – 20 ppm | 0.1 ppm |
| Desflurane | 0 – 20 ppm | 0.1 ppm |
| N2O | 0 – 100 ppm | 0.25 ppm |
| Isopropanol | 0 – 100 ppm | 0.5 ppm |
| Ethanol | 0 – 100 ppm | 0.5 ppm |
Options and accessories
How FTIR spectroscopy works
With FTIR (Fourier transform infrared) spectroscopy every compound absorbs infrared radiation at specific frequencies: analyzing the whole spectrum gives the concentration of every gas, all at the same time and continuously.
The ETG 9500 FTIR analyzer has a 0.5 cm⁻¹ resolution and a 5-metre optical path in a low-volume cell, for high selectivity and sensitivity, with detection limits well below 1 ppm. Automatic background acquisition, internal filters, cleaning valves and temperature and pressure control keep performance stable over time.