FTIR gas analyzer for versatile, uncompromising analysis
Multicomponent FTIR gas analyzer · industrial emissions monitoring · anaesthetic gas monitoring · greenhouse gas monitoring
ETG 9500 FTIR is the multicomponent FTIR gas analyzer designed by the in-house research and development team of ETG Risorse e Tecnologia. It measures many gases at the same time, continuously, from a single infrared spectrum, thanks to a high-resolution FTIR spectrometer with low interference and long-lasting components.

Three versions: 19" rack, portable and wall-mounted cabinet
The ETG 9500 FTIR analyzer comes in three formats, to suit the laboratory, field measurements and fixed plant installations.
ETG 9500 FTIR 19 – 19-inch rack
ETG 9500 FTIR P – portable FTIR analyzer
Portable analyzer for on-site measurements: mains-powered, with a 180 °C measuring cell for wet gases, or battery-powered, with a 60 °C cell and up to 10 hours of autonomy. IP54 optional.ETG 9500 FTIR WM – IP54 wall-mounted cabinet
ETG 9500 FTIR analyzer: key specifications
| Feature | Detail |
|---|---|
| Measuring principle | Fourier transform infrared (FTIR) spectroscopy |
| Optical path | 5 m as standard; other lengths on request |
| Response time (T90) | Less than 60 s |
| Measuring cell | 60 °C (applications without wet gases) or 180 °C (wet gases) |
| Connectivity | Ethernet and Wi-Fi as standard |
| Protection rating | WM: IP54 (higher grades on request) · P: IP54 optional |
| Ambient operating temperature | 5 to 45 °C |
| Autonomy (P battery version) | Up to 10 hours |
The measured gases depend on the configuration chosen for the application; other gases and combinations are available on request. For gas-specific data, contact ETG.
How FTIR spectroscopy works
With Fourier transform infrared spectroscopy (FTIR) every compound absorbs infrared radiation at specific frequencies: analyzing the spectrum with algorithms and mathematical models gives the concentration of each gas.
ETG adopted a 5-metre optical path in a low-volume cell, for high sensitivity and fast response (T90 below 60 seconds). Background acquisition, needed to compare the analyzed sample, is an easy, quick and automatic process, and makes it possible to exclude unwanted compounds whose absorption frequencies are very close to those of interest.
Applications of the ETG 9500 FTIR gas analyzer
Industrial emissions monitoring
Need a single compound? See the ETG 6900 analyzer.
Anaesthetic gas monitoring
Greenhouse gas monitoring
Many other applications
Composting process monitoring, catalyst control, CO2 quality control, gas purity analysis, toxicity monitoring after fire incidents, leak checks in chemical plants, research in biogas and syngas, volatile organic compounds analysis: FTIR technology makes the analyzer suitable for any use.