Multipoint anesthetic gas monitoring system for operating rooms
FTIR anesthetic gas monitor · cabinet · up to 48 sampling points · continuous monitoring · sevoflurane, isoflurane, desflurane, nitrous oxide

ETG 9500 Anesthetic Gases Monitor is the cabinet system for continuous monitoring of anesthetic gases in operating rooms and medical and laboratory environments. It includes a high-resolution FTIR analyzer, a multipoint sampling module and the supervision software, and detects sevoflurane, isoflurane, desflurane, nitrous oxide, carbon dioxide and water vapor, with optional configurations for additional compounds on request.
It serves any number of sampling points from 1 to 48, selected automatically and in sequence, with PTFE sampling lines up to 150 m per point. The cabinet can be equipped, on request, with an air conditioner chosen for the ambient conditions of the installation room.
Built for continuous, multi-room monitoring
Multipoint sampling
Continuous monitoring
Supervision software
Applications of the ETG 9500 Anesthetic Gases Monitor
Continuous monitoring of residual anesthetic gases in operating rooms, recovery rooms and laboratories: see the anesthetic gas monitoring application. Gas outlets are vented outside the installation room through a dedicated exhaust line, provided by the customer.
FTIR advantages over NDIR, gas chromatography and photoacoustic systems
Anesthetic gas monitoring has historically relied on NDIR analyzers, gas chromatography (GC) and, more recently, photoacoustic spectroscopy (PAS) with an infrared source. NDIR and PAS systems are generally based on a dedicated optical filter for each measured gas; the FTIR analyzer acquires the entire infrared spectrum and derives the concentration of all compounds of interest at once.
One analyzer, many gases
Faster, no carrier gas
Constant time, no alcohol interference
ETG 9500 Anesthetic Gases Monitor: key specifications
| Feature | Detail |
|---|---|
| Number of sampling points | Up to 48 (any number from 1 to 48) |
| Sampling lines | PTFE 6/4 mm, up to 150 m per point; quick fittings for 6 mm tubing |
| Pumps | Sample pump and fast loop pump (max 10 and 30 L/min) |
| Sample gas flow | 1 to 3 L/min |
| Power supply | 230 VAC 50/60 Hz |
| Power consumption | Max 800 W · typical 300 W |
| Connectivity | Ethernet (IP65), Wi-Fi, USB |
| Communication | Industrial protocols (Modbus, Profibus, Profinet…) (optional): SCADA integration with bidirectional communication (control from SCADA possible) |
| Alarm and analog outputs | 4-20 mA analog outputs (optional); 4 digital alarm outputs (optional). Status and alarm signals are also available via the communication protocol |
| Dimensions (L × W × D) | 600 × 1200 × 800 mm (reference values; vary with configuration) |
| Weight | 110 kg (reference value; varies with configuration) |
| Rating | IP 55, referring to the cabinet |
Operating conditions
| Feature | Detail |
|---|---|
| Sample gas pressure | Maximum 3 bar (abs.) |
| Sample gas temperature | 15 °C to 60 °C |
| Measuring cell temperature | 60 °C: the sample gas is brought to this temperature before analysis |
| Ambient conditions | Indoor use, up to 2000 m altitude. 5 °C to 40 °C; relative humidity 20% to 80% up to 31 °C, decreasing to 50% at 40 °C, non-condensing |
| Warm-up time | 30 minutes |
| Noise level | 65 dB(A) with air conditioner running; 30 dB(A) in configurations without air conditioner |
Analytical performance
| Feature | Detail |
|---|---|
| Zero drift | < 1% F.S. for every 2 weeks elapsed since the background |
| Linearity deviation | < 1% F.S. |
| Repeatability | ≤ 0.5% of the measured value |
| Temperature-induced drift | < 1% F.S. per 10 K of temperature change; ambient temperature changes are measured and compensated |
| Pressure influence | 1% F.S. for every 1% change in sample gas pressure; ambient pressure changes are measured and compensated |
| Response time (T90) | 30 seconds per sampling point |
| Background | Every 2 weeks (recommended), with nitrogen 5.0 connected only during background acquisition: a permanent nitrogen line is not required |
| Full cycle time | T90 × number of configured points: 24 minutes with 48 points, 12 minutes with 24 points |
Measured gases
| Measured gas | Range | LDL (lower detection limit) |
|---|---|---|
| H2O | 0 – 5 % | 0.05 % |
| CO2 | 0 – 2000 ppm | 10 ppm |
| Sevoflurane | 0 – 20 ppm | 0.1 ppm |
| Isoflurane | 0 – 20 ppm | 0.1 ppm |
| Desflurane | 0 – 20 ppm | 0.1 ppm |
| N2O | 0 – 100 ppm | 0.25 ppm |
| Isopropanol | 0 – 100 ppm | 0.5 ppm |
| Ethanol | 0 – 100 ppm | 0.5 ppm |
Options and accessories
How FTIR spectroscopy works
With FTIR (Fourier transform infrared) spectroscopy every compound absorbs infrared radiation at specific frequencies: analyzing the whole spectrum gives the concentration of every gas, all at the same time and continuously.
The ETG 9500 FTIR analyzer has a 0.5 cm⁻¹ resolution and a 5-metre optical path in a low-volume cell, for high selectivity and sensitivity, with detection limits well below 1 ppm. Automatic background acquisition, internal filters, cleaning valves and temperature and pressure control keep performance stable over time.