Portable TDL laser gas analyzer for on-site measurements
Portable TDL laser gas analyzer · mains-powered · rugged case · field measurements · ppm-level detection

ETG 6900 P is the portable version of the TDL laser gas analyzer developed by ETG Risorse e Tecnologia: a rugged case for selective measurements of specific gases, down to ppm levels, wherever the plant or the site requires them.
It is mains-powered, with a 7" touch screen, .csv data logging and connection via Ethernet or Wi-Fi. The heated-cell (-H) configuration, with heated line and optional heated probe, makes it suitable for hot, wet gases.
Built for measurements on site
Portable
Hot and wet gases
Connected
Applications of the ETG 6900 P gas analyzer
Industrial emissions monitoring
Greenhouse gas monitoring
Fugitive methane emissions
Many other applications
Leak checks, gas purity control, CO2 quality control, process monitoring and research: the laser sensor is chosen for the compound to be measured.
ETG 6900 P: key specifications
| Feature | Detail |
|---|---|
| Power supply | Mains: 230 VAC 50 Hz (other mains voltages/frequencies on request) |
| Power consumption | 70 W (non-heated cell) / 170 W (heated cell) · heated line 80 W/m · heated probe 80 W |
| Display / connectivity | 7" touch screen; Ethernet, Wi-Fi, USB |
| Communication protocol | 4-20 mA analog outputs (optional); 4 digital alarm outputs (optional); industrial protocols (Modbus, Profibus, Profinet…) (optional): SCADA integration with bidirectional communication (control from SCADA possible) |
| Data availability | .csv format, downloadable via USB or remote folder (SMB) |
| Dimensions (L × W × D) / weight | 625 × 500 × 297 mm · 7.2 kg |
| Rating | IP54 (optional) |
| Certification | CE |
| Sample pump | Optional (heated pump with the heated cell) |
| Sample flow / fittings | 0.1 to 3 l/min (non-heated cell); 1.5 to 5 l/min (heated cell) · for 6 mm O.D. tubing |
| Laser sensors | One laser sensor, selected for the gases to be measured |
Operating conditions
| Feature | Detail |
|---|---|
| Sample gas | 800-1100 mbar; ambient temperature (non-heated cell) or up to 190 °C (heated cell, -H) |
| Sample gas characteristics | Oil-free, dust-free; non-condensing (non-heated cell) or up to 30% vol H2O (heated cell) |
| Measuring cell temperature | Not heated; 190 °C with heated cell (-H), up to 220 °C |
| Ambient conditions | 0 to 50 °C; ≤ 95% RH non-condensing; -300 to 3000 m |
| Calibration | Zero with N2; span with a cylinder of the target gas |
| Warm-up time | 30 s to start; usable after about 3 min; stable after about 30 min (heated cell: about 30 min to reach 190 °C, fully stable after about 60 min) |
Analytical performance
| Feature | Detail |
|---|---|
| Measuring principle | TDLAS (tunable diode laser absorption spectroscopy) |
| Accuracy | ±2% of full scale (or ±2 ppm / 2% of reading, whichever is greater) |
| CH4 | 0-100 ppm up to 0-40,000 ppm · precision ≤ 0.8 ppm or 0.6% of reading · T90 ≤ 1.8 s |
| CH4 + CO2 | CH4 0-100 ppm, CO2 0-10% vol · precision CH4 ≤ 0.8 ppm, CO2 ≤ 2,000 ppm · T90 ≤ 30 s (0.3 l/min) |
| CH4 + C2H6 | CH4 0-100 ppm up to 0-40,000 ppm, C2H6 0-1,000 ppm · T90 CH4 ≤ 1.8 s, C2H6 ≤ 15 s |
| NH3 + H2O | NH3 0-100 ppm, H2O 0-5% vol · precision NH3 ≤ 0.8 ppm, H2O ≤ 0.2% vol · T90 ≤ 3 s |
| C2H2 + H2O | C2H2 0-100 ppm (linear up to 500 ppm), H2O 0-10% vol · T90 60 s (0.6 l/min) |
| H2O (-H) | 0-30% vol · accuracy ±0.5% vol · precision 0.2% vol (1 s) / 0.1% vol (10 s) · T90 2 s (3 l/min) |
| NH3 + H2O (-H) | NH3 0-100 ppm (min. range 0-20 ppm), H2O 0-30% vol · NH3 precision 1.8 ppm (1 s) / 0.8 ppm (10 s) · T90 2 s (3 l/min) |
| HCl + H2O (-H) | HCl 0-50 ppm (linear up to 500 ppm, no precision guarantee), H2O 0-30% vol · detection limit < 0.4 ppm, precision < 0.6 ppm (10 s avg.), accuracy ±2 ppm · T90 20 s (3 l/min) |
Ranges, precision and response times as specified for each laser sensor. The heated-cell (-H) configuration includes a heated measuring cell (190 °C), a heated sample pump (if a pump is installed) and a heated sample line (mandatory).
Options and accessories
How TDL laser spectroscopy works
TDLAS (tunable diode laser absorption spectroscopy) tunes a diode laser across a very narrow absorption line of the target gas: the laser line is much narrower than the absorption line of the gas, so the resolution is much higher than with other technologies, and the analyzer detects specific compounds selectively, in real time and with minimal interference, even at ppm concentrations.
The laser sensor is chosen for the gases to be measured. TDL technology does not require frequent calibrations, which keeps maintenance and running costs low.
Three formats, one TDL laser technology
The ETG 6900 uses the same laser technology in every format: pick the one that fits your application.
ETG 6900 19" – rack
ETG 6900 WM – wall-mounted
Need to measure many gases at the same time? See the ETG 9500 FTIR multicomponent analyzer.