19-inch rack TDL laser gas analyzer for selective, continuous measurement
TDL laser gas analyzer · 19" 4U rack · continuous measurement · ppm-level detection · heated cell for hot, wet gases

ETG 6900 19" is the rack version of the TDL laser gas analyzer developed by ETG Risorse e Tecnologia. A tunable diode laser detects a specific gas selectively, in real time and down to ppm levels, with very fast response and little maintenance.
Housed in a standard 19" 4U rack enclosure with a 7" touch screen, it is made for fixed installations. Data are logged in .csv, with remote access via Ethernet or Wi-Fi. The laser sensor is selected for the gases you need to measure.
Built for fixed installations and continuous measurement
Selective
Rack-mounted
Connected
Applications of the ETG 6900 19" gas analyzer
Industrial emissions monitoring
Greenhouse gas monitoring
Fugitive methane emissions
Many other applications
Leak checks, gas purity control, CO2 quality control, process monitoring and research: the laser sensor is chosen for the compound to be measured.
ETG 6900 19": key specifications
| Feature | Detail |
|---|---|
| Power supply | 230 VAC 50 Hz (other mains voltages/frequencies on request) |
| Power consumption | 70 W (non-heated cell) / 170 W (heated cell) · heated line 80 W/m · heated probe 80 W |
| Display / connectivity | 7" touch screen; Ethernet, Wi-Fi, USB |
| Communication protocol | 4-20 mA analog outputs (optional); 4 digital alarm outputs (optional); industrial protocols (Modbus, Profibus, Profinet…) (optional): SCADA integration with bidirectional communication (control from SCADA possible) |
| Data availability | .csv format, downloadable via USB or remote folder (SMB) |
| Dimensions (L × W × D) / weight | 483 × 177 × 550 mm (19" rack, 4U) · 11 kg |
| Certification | CE |
| Sample pump | Optional (heated pump with the heated cell) |
| Sample flow / fittings | 0.1 to 3 l/min (non-heated cell); 1.5 to 5 l/min (heated cell) · for 6 mm O.D. tubing |
| Laser sensors | One laser sensor, selected for the gases to be measured |
Operating conditions
| Feature | Detail |
|---|---|
| Sample gas | 800-1100 mbar; ambient temperature (non-heated cell) or up to 190 °C (heated cell, -H) |
| Sample gas characteristics | Oil-free, dust-free; non-condensing (non-heated cell) or up to 30% vol H2O (heated cell) |
| Measuring cell temperature | Not heated; 190 °C with heated cell (-H), up to 220 °C |
| Ambient conditions | 0 to 50 °C; ≤ 95% RH non-condensing; -300 to 3000 m |
| Calibration | Zero with N2; span with a cylinder of the target gas |
| Warm-up time | 30 s to start; usable after about 3 min; stable after about 30 min (heated cell: about 30 min to reach 190 °C, fully stable after about 60 min) |
Analytical performance
| Feature | Detail |
|---|---|
| Measuring principle | TDLAS (tunable diode laser absorption spectroscopy) |
| Accuracy | ±2% of full scale (or ±2 ppm / 2% of reading, whichever is greater) |
| CH4 | 0-100 ppm up to 0-40,000 ppm · precision ≤ 0.8 ppm or 0.6% of reading · T90 ≤ 1.8 s |
| CH4 + CO2 | CH4 0-100 ppm, CO2 0-10% vol · precision CH4 ≤ 0.8 ppm, CO2 ≤ 2,000 ppm · T90 ≤ 30 s (0.3 l/min) |
| CH4 + C2H6 | CH4 0-100 ppm up to 0-40,000 ppm, C2H6 0-1,000 ppm · T90 CH4 ≤ 1.8 s, C2H6 ≤ 15 s |
| NH3 + H2O | NH3 0-100 ppm, H2O 0-5% vol · precision NH3 ≤ 0.8 ppm, H2O ≤ 0.2% vol · T90 ≤ 3 s |
| C2H2 + H2O | C2H2 0-100 ppm (linear up to 500 ppm), H2O 0-10% vol · T90 60 s (0.6 l/min) |
| H2O (-H) | 0-30% vol · accuracy ±0.5% vol · precision 0.2% vol (1 s) / 0.1% vol (10 s) · T90 2 s (3 l/min) |
| NH3 + H2O (-H) | NH3 0-100 ppm (min. range 0-20 ppm), H2O 0-30% vol · NH3 precision 1.8 ppm (1 s) / 0.8 ppm (10 s) · T90 2 s (3 l/min) |
| HCl + H2O (-H) | HCl 0-50 ppm (linear up to 500 ppm, no precision guarantee), H2O 0-30% vol · detection limit < 0.4 ppm, precision < 0.6 ppm (10 s avg.), accuracy ±2 ppm · T90 20 s (3 l/min) |
Ranges, precision and response times as specified for each laser sensor. The heated-cell (-H) configuration includes a heated measuring cell (190 °C), a heated sample pump (if a pump is installed) and a heated sample line (mandatory).
Options and accessories
How TDL laser spectroscopy works
TDLAS (tunable diode laser absorption spectroscopy) tunes a diode laser across a very narrow absorption line of the target gas: the laser line is much narrower than the absorption line of the gas, so the resolution is much higher than with other technologies, and the analyzer detects specific compounds selectively, in real time and with minimal interference, even at ppm concentrations.
The laser sensor is chosen for the gases to be measured. TDL technology does not require frequent calibrations, which keeps maintenance and running costs low.
Three formats, one TDL laser technology
The ETG 6900 uses the same laser technology in every format: pick the one that fits your application.
ETG 6900 P – portable
ETG 6900 WM – wall-mounted
Need to measure many gases at the same time? See the ETG 9500 FTIR multicomponent analyzer.