Wall-mount syngas analyzer for gasification and pyrolysis, with integrated gas cleaning
Methane (CH4) · Carbon dioxide (CO2) · Carbon monoxide (CO) · Hydrogen (H2) · Oxygen (O2) · Syngas analyzer · Gasifiers, pyrolysis and furnaces

ETG 6500 SYN is a stainless-steel wall-mount syngas analyzer for continuous monitoring of gas from gasification and pyrolysis. NDIR measures CH4, CO2 and CO, a TCD measures H2 and a long-life electrochemical cell measures O2, while N2 and calorific value are calculated continuously.
Plug&Play and tar-ready: a Peltier chiller and a gas cleaning system equivalent to the ETG PSS100 are built into the cabinet.
Built for raw syngas, 24/7
Integrated sample cleaning
Heating value in real time
Expandable and connected
Where the ETG 6500 SYN is used
Gasification and pyrolysis
Metallurgical and heat-treatment furnaces
Research centers and system integrators
ETG 6500 SYN: key specifications
| Feature | Detail |
|---|---|
| Power supply | 230 VAC 50 Hz (other mains voltages/frequencies on request) |
| Power consumption | Max 300 W · typical 150 W |
| Display | 7" touch screen |
| Connectivity | Ethernet, USB, Wi-Fi |
| Communication protocol | 4-20 mA (standard when no industrial protocol is installed); Modbus, Profibus, Profinet (optional) |
| Data availability | .csv via USB or remote folder (SMB) |
| Dimensions / weight | 600 × 800 × 350 mm · 35 kg |
| Rating / certification | IP54 · CE |
| Sample conditioning | Peltier chiller (standard) · integrated gas cleaning for tar, dust and water, equivalent to the ETG PSS100 |
| Sample pump | Optional (internal) |
| Sample flow / fittings | 0.5–1.5 l/min · Swagelok, 6 mm O.D. |
| Sample gas | 0.1–0.3 bar · 5–40 °C non-condensing |
| Operating conditions | 0–50 °C · up to 95% RH non-condensing · −300 to 3000 m |
| Warm-up | 30 s ready · 3 min usable · 30 min full performance |
| Calibration | Manual zero and span (automatic zero with ambient air optional) |
Analytical performance
| Gas | Principle | Range | Response |
|---|---|---|---|
| Methane (CH4) | NDIR | 0 – 100% vol | T90 & T10 < 10 s |
| Carbon dioxide (CO2) | NDIR | 0 – 100% vol | T90 & T10 < 10 s |
| Carbon monoxide (CO) | NDIR | 0 – 100% vol | T90 & T10 < 10 s |
| Hydrogen (H2) | TCD, automatically corrected | 0 – 10 / 20 / 50 / 100% vol | T90 & T10 < 20 s |
| Oxygen (O2) | Electrochemical (ECD) | 0 – 25% vol | < 30 s |
| Hydrocarbons (CnHm) – optional | NDIR | 0 – 5 / 10 / 20% vol | T90 & T10 < 10 s |
| Hydrogen sulphide (H2S) – optional | NDUV | Ranges on request | On request |
| Ammonia (NH3) – optional | ECD | 0 – 200 / 1000 ppm | On request |
Resolution 0.01%. Accuracy ±1% F.S. for CH4, CO2, CO and CnHm, ±2% F.S. for H2 and O2. N2 and calorific value are calculated continuously with a proprietary algorithm (standard). Other gases (including paramagnetic O2) and other ranges are available on request. O2 response: < 30 s from ambient air to 0.15% O2.
ETG software: touch interface, trends and remote access
A 7" touch interface shows live values, trends and alarms. Data are logged in .csv and the display can be viewed remotely via VNC. Analyzers fitted with the optional Modbus, Profibus or Profinet protocols integrate with SCADA systems and offer two-way communication, so they can also be controlled from SCADA when needed.


Options and accessories
Three formats for syngas analysis
ETG syngas analyzers share the same measurement technology, the same software and the same performance, and differ in format: pick the one that fits your plant or your measurement campaign.
ETG MCA100 SYN 19" – rack
ETG MCA100 SYN P – portable
Syngas with tar, dust or water vapour? Pair the analyzer with the ETG PSS100 sample treatment system.