FTIR anesthetic gas monitors for operating rooms: multipoint cabinet and portable analyzer
FTIR anesthetic gas monitoring · multi-gas analysis from a single spectrum · up to 48 sampling points · portable analyzer · software and algorithms developed by ETG

ETG 9500 anesthetic gas monitors are designed for the environmental monitoring of residual anesthetic gases in operating rooms and medical and laboratory environments. They are based on the ETG 9500 FTIR analyzer, integrated by ETG with software and algorithms developed by its in-house research and development team, and measure sevoflurane, isoflurane, desflurane, nitrous oxide, carbon dioxide and water vapor at the same time, from a single infrared spectrum.
Two systems share the same FTIR technology: a cabinet system with multipoint sampling for continuous monitoring of up to 48 points, and a portable analyzer for spot checks, with an optional multipoint case.
Two ETG 9500 systems for anesthetic gas monitoring

ETG 9500 Anesthetic Gases Monitor

ETG 9500 P Anesthetic Gases Monitor
Why monitor anesthetic gases continuously
Compared to periodic spot checks, a continuous monitoring system makes it possible to:
- Detect both average exposure values (TWA) and peak values, both used as reference by industry guidelines for assessing the risk of exposure to anesthetic gases
- Build a historical data record useful for audits, inspections and reporting
- Trigger automatic real-time alarms whenever configured thresholds are exceeded
- Verify, under real operating conditions, the effectiveness of air exchange and waste anesthetic gas evacuation (scavenging) systems, promptly identifying leaks or ventilation anomalies
The portable analyzer is the right tool for periodic exposure checks, verification after a suspected leak or an equipment change, and rooms without a fixed monitoring system.
How FTIR spectroscopy works
With FTIR (Fourier transform infrared) spectroscopy every compound absorbs infrared radiation at specific frequencies: analyzing the whole spectrum gives the concentration of every gas, all at the same time and continuously.
The ETG 9500 FTIR analyzer has a 0.5 cm⁻¹ resolution and a 5-metre optical path in a low-volume cell, for high selectivity and sensitivity, with detection limits well below 1 ppm. Automatic background acquisition, internal filters, cleaning valves and temperature and pressure control keep performance stable over time.
Advantages of FTIR over NDIR, gas chromatography and photoacoustic spectroscopy
Anesthetic gas monitoring has historically relied on NDIR analyzers, gas chromatography (GC) and, more recently, photoacoustic spectroscopy (PAS) with an infrared source. NDIR and PAS systems are generally based on a dedicated optical filter for each measured gas; the FTIR analyzer acquires the entire infrared spectrum and derives the concentration of all compounds of interest at once.
One analyzer, many gases
Faster, no carrier gas
Constant time, no alcohol interference
- New gases or different measuring ranges are added by software, even remotely over a LAN, with no hardware change
- The full spectrum is stored: historical data can be re-analyzed later, even for compounds not originally configured
- Sub-ppm detection limits, kept in the presence of humidity and carbon dioxide
Application: anesthetic gas monitoring in operating rooms
Operating rooms, recovery rooms and any room where anesthetic gases are used: see the anesthetic gas monitoring application for details.